Clinical dose assessment for scanned carbon-ion radiotherapy using linear energy transfer measurements and Monte Carlo simulations

Author:

Nakaji TakuORCID,Kanai Tatsuaki,Takashina Masaaki,Matsumura Akihiko,Osaki Kohei,Yagi MasashiORCID,Tsubouchi Toshiro,Hamatani Noriaki,Ogawa Kazuhiko

Abstract

Abstract Objective. Dosimetric commissioning of treatment planning systems (TPS) focuses on validating the agreement of the physical dose with experimental data. For carbon-ion radiotherapy, the commissioning of the relative biological effectiveness (RBE) is necessary to predict the clinical outcome based on the radiation quality of the mixed radiation field. In this study, we proposed a approach for RBE commissioning using Monte Carlo (MC) simulations, which was further strengthen by RBE validation based on linear energy transfer (LET) measurements. Approach. First, we tuned the MC simulation based on the results of dosimetric experiments including the beam ranges, beam sizes, and MU calibrations. Furthermore, we compared simulated results to measured depth- and radial-LET distributions of the 430 MeV u−1 carbon-ion spot beam with a 1.5 mm2, 36 μm thick silicon detector. The measured dose-averaged LET (LETd) and RBE were compared with the simulated results. The RBE was calculated based on the mixed beam model with linear-quadratic parameters depending on the LET. Finally, TPS-calculated clinical dose profiles were validated through the tuned MC-based calculations. Main results. A 10 keV μm−1 and 0.15 agreement for LETd and RBE, respectively, were found between simulation and measurement results obtained for a 2σ lateral size of 430 MeV u−1 carbon-ion spot beam in water. These results suggested that the tuned MC simulation can be used with acceptable precision for the RBE and LET calculations of carbon-ion spot beam within the clinical energy range. For physical and clinical doses, the TPS- and MC-based calculations showed good agreements within 1.0% at the centre of the spread-out Bragg peaks. Significance. The tuned MC simulation can accurately reproduce the actual carbon-ion beams, and it can be used to validate the physical and clinical dose distributions calculated by TPS. Moreover, the MC simulation can be used for dosimetric commissioning, including clinical doses, without LET measurements.

Funder

Japan Society for the Promotion of Science

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

Reference34 articles.

1. Geant4—a simulation toolkit;Agostinelli;Nucl. Instruments Methods Phys. Res. A,2003

2. The PTSim and TOPAS Projects, Bringing Geant4 to the particle therapy clinic;Akagi;Prog. Nucl. Sci. Technol.,2011

3. Geant4 developments and applications;Allison;IEEE Trans. Nucl. Sci.,2006

4. Recent developments in Geant4;Allison;Nucl. Instruments Methods Phys. Res. A,2016

5. GEANT4 based simulation framework for particle therapy system;Aso,2007