Investigation of test methods for QC in dual-energy based contrast-enhanced digital mammography systems: II. Artefacts/uniformity, exposure time and phantom-based dosimetry

Author:

Marshall N WORCID,Cockmartin L,Bosmans H

Abstract

Abstract Part II of this study describes constancy tests for artefacts and image uniformity, exposure time, and phantom-based dosimetry; these are applied to four mammography systems equipped with contrast enhanced mammography (CEM) capability. Artefacts were tested using a breast phantom that simulated breast shape and thickness change at the breast edge. Image uniformity was assessed using rectangular poly(methyl)methacrylate PMMA plates at phantom thicknesses of 20, 40 and 60 mm, for the low energy (LE), high energy (HE) images and the recombined CEM image. Uniformity of signal and of the signal to noise ratio was quantified. To estimate CEM exposure times, breast simulating blocks were imaged in automatic exposure mode. The resulting x-ray technique factors were then set manually and exposure time for LE and HE images and total CEM acquisition time was measured with a multimeter. Mean glandular dose (MGD) was assessed as a function of simulated breast thickness using three different phantom compositions: (i) glandular and adipose breast tissue simulating blocks combined to give glandularity values that were typical of those in a screening population, as thickness was changed (ii) PMMA sheets combined with polyethylene blocks (iii) PMMA sheets with spacers. Image uniformity was superior for LE compared to HE images. Two systems did not generate recombined images for the uniformity test when the detector was fully covered. Acquisition time for a CEM image pair for a 60 mm thick breast equivalent phantom ranged from 3.4 to 10.3 s. Phantom composition did not have a strong influence on MGD, with differences generally smaller than 10%. MGD for the HE images was lower than for the LE images, by a factor of between 1.3 and 4.0, depending on system and simulated breast thickness. When combined with the iodine signal assessment in part I, these tests provide a comprehensive assessment of CEM system imaging performance.

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3