Abstract
Abstract
We have developed a scalable detector readout design for a 100 ps coincidence time resolution (CTR) time of flight (TOF) positron emission tomography (PET) detector technology. The basic scintillation detectors studied in this paper are based on 2 × 4 arrays of 3 × 3 × 10 mm3 ‘fast-LGSO:Ce’ scintillation crystals side-coupled to 6 × 4 arrays of 3 × 3 mm2 silicon photomultipliers (SiPMs). We employed a novel mixed-signal front-end electronic configuration and a low timing jitter Field Programming Gate Array-based time to digital converter for data acquisition. Using a 22Na point source, >10 000 coincidence events were experimentally acquired for several SiPM bias voltages, leading edge time-pickoff thresholds, and timing channels. CTR of 102.03 ± 1.9 ps full-width-at-half-maximum (FWHM) was achieved using single 3 × 3 × 10 mm3 ‘fast-LGSO’ crystal elements, wrapped in Teflon tape and side coupled to a linear array of 3 SiPMs. In addition, the measured average CTR was 113.4 ± 0.7 ps for the side-coupled 2 × 4 crystal array. The readout architecture presented in this work is designed to be scalable to large area module detectors with a goal to create the first TOF-PET system with 100 ps FWHM CTR.
Funder
National Institutes of Health
National Research Foundation of Korea
Subject
Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology
Cited by
23 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献