Abstract
Abstract
Objective. Time-of-flight (TOF) scatter rejection allows for identifying and discarding scattered photons without the use of an anti-scatter grid (ASG). Although TOF scatter rejection was initially presented for cone-beam computed tomography, we propose, herein, to extend this approach to x-ray radiography. This work aims to evaluate with simulations if TOF scatter rejection can outperform ASGs for radiography. Approach. GATE was used to simulate the radiography of a head and a torso and a water cylinder with bone inserts in a system with total timing jitters from 0 ps up to 500 ps full-width-at-half-maximum. The transmission factor of TOF scatter rejection for primary and scattered photons was evaluated as if it were a virtual ASG. Main results. With a total timing jitter of 50 ps, TOF scatter rejection can reach a selectivity of 4.93 with a primary photons transmission of 99%. Reducing the timing jitter close to 0 ps increases the selectivity up to 15.85 for a head and torso radiography, outperforming typical ASGs which usually have a selectivity from 2.5 to 10 with a primary photons transmission from 50% to 70%. Significance. This suggests that TOF scatter rejection may be suitable to replace ASGs in applications requiring lower radiation exposure if sufficiently low timing jitter is achieved.