Characterization of real-time cine MR imaging distortion on 0.35 T MRgRT with concentric cine imaging QA phantom

Author:

Marasini ShantiORCID,Cole Mike,Curcuru Austen,Dyke Lara M,Gach H Michael,Flores Rocco,Kim TaehoORCID

Abstract

Abstract Objective. Real-time MRgRT uses 2D-cine imaging for target tracking and motion evaluation. Rotation of gantry induced B 0 off-resonance, resulting in image artifacts and imaging isocenter-shift precluding MR-guided arc therapy. Standard MRI phantoms designed for higher resolution images face challenges when low-resolution cine imaging is needed to achieve high frame rates. This work aimed to examine the spatial accuracy including geometric distortion and isocenter shift in real-time during gantry rotation on a 0.35 T MR-Linac using the concentric Cine imaging quality assurance (QA) phantom and its associated image analysis software. Approach. The Cine imaging QA phantom consists of two concentric shells of low-T1 mineral oil and a central alignment structure. The phantom was scanned on three different MRI systems; 0.55 T Siemens Free.Max, 1.5 T Philips Ingenia, and 0.35 T ViewRay MRIdian MR-Linac using 2D balanced steady-state free precession (bSSFP) imaging sequence. In addition, bSSFP cine MRI with the banding artifact correction was tested on 0.35 T ViewRay MR-Linac. Images from the MR-Linac were acquired with the Linac gantry stationary and rotating from gantry 300°→ 0° and vice versa. Three orthogonal image planes were scanned excluding the 1.5 T Philips Ingenia, where only the axial plane was scanned. The image analysis software calculated the distortion values as well as the isocenter position for each cine frame. Main results. The geometric distortion of cine imaging on MRIs and MR-Linac at gantry stationary are within 1 mm while the substantial geometric distortion of 2 and 2.2 mm were observed on 0.35 T MR-Linac while rotating the gantry clockwise (300°→ 0°) and counterclockwise 0°→ 300° respectively. The average imaging isocenter shift was 0.1 mm for both MRIs and the static gantry and imaging isocenter shift of ≤1.5 mm was observed during the gantry rotation. The imaging isocenter shift decreased by 1 ± 0.2 mm clockwise and counterclockwise with B 0 compensation. Significance. The concentric Cine imaging QA phantom and its associated software effectively demonstrate the image distortion on real-time cine imaging on regular MRIs and 0.35 T MR-Linac. The results of significant geometric distortion with a rotating gantry in the MR-Linac system require further investigation to alleviate the extent of the image distortion.

Publisher

IOP Publishing

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3