A dedicated phantom design for positron emission mammography performance evaluation

Author:

Torres-Urzúa Luis Fernando,Alva-Sánchez HéctorORCID,Martínez-Dávalos ArnulfoORCID,García-Pérez Francisco Osvaldo,Peruyero-Rivas Rocío Marlene,Rodríguez-Villafuerte MercedesORCID

Abstract

Abstract A standard protocol for performance evaluation of positron emission mammography (PEM) systems has not yet been established. In this work we propose a methodology based on the design of specific phantoms for this imaging modality with component dimensions in accordance with typical breast lesion sizes together with the adaptation of current international protocols designed for clinical and preclinical positron emission tomographs (PET) systems. This methodology was used to evaluate the performance of the Flex Solo II PEM scanner in terms of spatial resolution, uniformity and contrast lesion detectability, recovery coefficients and spill-over ratios. Positron range effects were studied with 18F and 68Ga, which have very different energy spectra. Our results indicate that in-plane spatial resolution of the system is around 3.0 mm and 4.4 mm for 18F and 68Ga, respectively. Lesion detectability tests with sphere diameters between 4 and 10 mm confirmed that the PEM system can resolve all the spheres (hot or cold). Percent contrast values for 18F lie between 6%–38% and 34%–51% for hot- and cold- spheres, respectively; the corresponding intervals for 68Ga are lower, 4%–25% and 32%–44%. Regarding uniformity quantification, the system shows percentage standard deviations within 4.9%–5.7%, while the percent background variability measurements ranged between 6.7% and 10.9% for both radionuclides. Recovery coefficients measured with hot rod diameters between 1.5 and 9 mm, have values between 0.2–1.05 and 0.17–0.69 for 18F and 68Ga, respectively. Spill-over ratios have large values (0.22 in average) for both radionuclides. Our results indicate that the phantoms and the methodology developed in this work can serve as the basis for establishing an image quality protocol for the systematic evaluation of PEM systems, with a potential extension for performance evaluation of dedicated breastPET scanners.

Funder

Consejo Nacional de Ciencia y Tecnología

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Quantitative analysis of the effect of the magnetic field generated by a PET/MR scanner on positron range;Physics in Medicine & Biology;2024-09-09

2. Crystal scatter effects in a large-area dual-panel Positron Emission Mammography system;PLOS ONE;2024-03-01

3. Experimental studies of the effect of the magnetic field in PET/MR on positron range;2023 IEEE Nuclear Science Symposium, Medical Imaging Conference and International Symposium on Room-Temperature Semiconductor Detectors (NSS MIC RTSD);2023-11-04

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3