Novel framework for determining TPS-calculated doses corresponding to detector locations using 3D camera in in vivo surface dosimetry

Author:

Sheen HeesoonORCID,Park Ye-in,Cho Min-Seok,Son Junyoung,Shin Han-Back,Han Min CheolORCID,Kim Hojin,Lee HoORCID,Kim Dong Wook,Kim Jin Sung,Hong Chae-SeonORCID

Abstract

Abstract Purpose. To address the shortcomings of current procedures for evaluating the measured-to-planned dose agreement in in vivo dosimetry (IVD), this study aimed to develop an accurate and efficient novel framework to identify the detector location placed on a patient’s skin surface using a 3D camera and determine the planned dose at the same anatomical position corresponding to the detector location. Methods. Breast cancer treatment was simulated using an anthropomorphic adult female phantom (ATOM 702D; CIRS, Norfolk, VA, USA). An optically stimulated luminescent dosimeter was used for surface dose measurements (MyOSLchip, RadPro International GmbH, Germany) at six IVD points. Three-dimensional surface imaging (3DSI) of the phantom with the detector was performed in the treatment position using a 3D camera. The developed framework, iSMART, was designed to import 3DSI and treatment planning data for determining the position of the IVD detectors in the 3D treatment planning DICOM image. The clinical usefulness of iSMART was evaluated in terms of accuracy and efficiency, for comparison with the results obtained using cone-beam computed tomography (CBCT) image guidance. Results. The relative dose difference between the planned doses determined using iSMART and CBCT images displayed similar accuracies (within approximately ±2.0%) at all detector locations. The relative dose differences between the planned and measured doses at the six detector locations ranged from –4.8% to 3.1% for the CBCT images and –3.5% to 2.1% for iSMART. The total time required to read the planned doses at six detector locations averaged at 8.1 and 0.8 min for the CBCT images and iSMART, respectively. Conclusions. The proposed framework can improve the robustness of IVD analyses and aid in accurate and efficient evaluations of the measured-to-planned dose agreement.

Funder

National Research Foundation of Korea

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3