On the length used for CT ionization chambers to determine CTDI

Author:

DeWerd Larry A,Khan Ahtesham UllahORCID,Jensen Andrew R

Abstract

Abstract Objective. Computed tomography dose index (CTDI) calculations based on measurements made with CT ionization chambers require characterization of two chamber properties: radiation sensitivity and effective length. The sensitivity of a CT ionization chamber is currently determined in some countries by calibration in an x-ray field that irradiates the entire chamber. Determination of the effective length is left to the user, and this value is frequently assumed to be equivalent to the nominal length—typically 100 mm—stated by the manufacturer. This assumption undermines the intention and usefulness of CTDI calculation. Thus, a slit-based calibration, N KL, of the CT ionization chambers was proposed by collimating the x-ray beam to a well-defined aperture width. The aim of this work is to compare the two methods. Approach. Four different CT ionization chambers (Standard Imaging Exradin A101, Radcal 10x5-3CT, Victoreen 500-100, and Capintec PC-4P) are investigated in this work. Sensitivity profiles were measured for all four chambers and effective/rated chamber lengths were calculated. A novel Monte-Carlo based correction was proposed to account for the presence of the aperture. CTDI was calculated and compared for two calibration beams as well as for a commercial CT scanner using Exradin A101 and Radcal 10x5-3CT chambers. Main results. The nominal chamber length was found to deviate up to 21% compared to the effective length. Correction for the aperture depended on the aperture opening size. CTDI calculation results illustrate the potential 17% error in CTDI calculation that can be caused by assuming the effective chamber length is equivalent to the manufacturer’s stated nominal length. CTDI calculations with CT ionization chambers calibrated with an air-kerma length calibration method yield the smallest variation in the CTDI regardless of the chamber model. Significance. To avoid an erroneous CTDI, information regarding the chamber’s effective length must be included in the calibration or stated by the manufacturer. Alternatively, a slit-based calibration can be performed.

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

Reference29 articles.

1. Calibration of ionization chambers in air kerma length;Bochud;Phys. Med. Biol.,2001

2. Determination of CT scanner radiation output: correction factors for partial irradiation of thimble and pencil ionisation chambers by collimated fan beams;Cross;Australas. Phys. Eng. Sci. Med.,1998

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3