Visualization of gold fiducial markers in the prostate using phase-cycled bSSFP imaging for MRI-only radiotherapy

Author:

Shcherbakova YuliaORCID,Bartels Lambertus W,Mandija StefanoORCID,Beld EllisORCID,Seevinck Peter R,van der Voort van Zyp Jochem R N,Kerkmeijer Linda G W,Moonen Chrit T W,Lagendijk Jan J W,van den Berg Cornelis A T

Abstract

Abstract In this work, we present a new method for visualization of fiducial markers (FMs) in the prostate for MRI-only radiotherapy with a positive contrast directly at the MR console. The method is based on high bandwidth phase-cycled balanced steady-state free precession (bSSFP) sequence, which is available on many clinical scanners, does not require any additional post-processing or software, and has a higher signal-to-noise (SNR) compared to conventional gradient-echo (GE) imaging. Complex phase-cycled bSSFP data is acquired with different RF phase increment settings such that the manifestation of the artifacts around FMs in the acquired complex images is different for each dynamic acquisition and depends on the RF phase increment used. First, we performed numerical simulations to investigate the complex-valued phase-cycled bSSFP signal in the presence of a gold FM, and to investigate the relation of the true physical location of the FM with the geometrical manifestation of the artifacts. Next, to validate the simulations, we performed phantoms and in vivo studies and compared the experimentally obtained artifacts with those predicted in simulations. The accuracy of the method was assessed by comparing the distances between the FM’s centers and the center of mass of FMs system measured using phase-cycled bSSFP MR images and using reference CT (or MRI-only) images. The results show accurate (within 1 mm) matching of FMs localization between CT and MR images on five patients, proving the feasibility of in vivo FMs detection on MR images only. The FMs show a positive contrast with respect to the prostate background on real/imaginary phase-cycled bSSFP images, which was confirmed by simulations. The proposed method facilitates robust FMs visualization with positive contrast directly at the MR console, allowing RT technicians to obtain immediate feedback on the anticipated feasibility of accurate FMs localization while the patient is being scanned.

Funder

The Netherlands Organisation for Scientific Research (NWO), Domain Applied and Engineering Sciences

Publisher

IOP Publishing

Subject

Radiology Nuclear Medicine and imaging,Radiological and Ultrasound Technology

Reference39 articles.

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3