Monte Carlo-calculated beam quality and perturbation correction factors validated against experiments for Farmer and Markus type ionization chambers in therapeutic carbon-ion beams

Author:

Urago YukaORCID,Sakama Makoto,Sakata Dousatsu,Fukuda Shigekazu,Katayose Tetsurou,Chang WeishanORCID

Abstract

Abstract Objective. In current dosimetry protocols, the estimated uncertainty of the measured absorbed dose to water D w in carbon-ion beams is approximately 3%. This large uncertainty is mainly contributed by the standard uncertainty of the beam quality correction factor k Q. In this study, the k Q values in four cylindrical chambers and two plane-parallel chambers were calculated using Monte Carlo (MC) simulations in the plateau region. The chamber-specific perturbation correction factor P of each chamber was also determined through MC simulations. Approach. k Q for each chamber was calculated using MC code Geant4. The simulated k Q ratios in subjected chambers and reference chambers were validated through comparisons against our measured values. In the measurements in Heavy-Ion Medical Accelerator in Chiba, k Q ratios were obtained from D w values of 60Co, 290- and 400 MeV u−1 carbon-ion beams that were measured with the subjected ionization chamber and the reference chamber. In the simulations, f Q (the product of the water-to-air stopping power ratio and P) was acquired from D w and the absorbed dose to air calculated in the sensitive volume of each chamber. k Q values were then calculated from the simulated f Q and the literature-extracted W air and compared with previous publications. Main results. The calculated k Q ratios in the subjected chambers to the reference chamber agreed well with the measured k Q ratios. The k Q uncertainty was reduced from the current recommendation of approximately 3% to 1.7%. The P values were close to unity in the cylindrical chambers and nearly 1% above unity in the plane-parallel chambers. Significance. The k Q values of carbon-ion beams were accurately calculated in MC simulations and the k Q ratios were validated through ionization chamber measurements. The results indicate a need for updating the current recommendations, which assume a constant P of unity in carbon-ion beams, to recommendations that consider chamber-induced differences.

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

Reference40 articles.

1. GEANT4—a simulation toolkit;Agostinelli;Nucl. Instrum. Methods Phys. Res. A,2003

2. Absorbed dose determination in external beam radiotherapy: an international code of practice for dosimetry based on standards of absorbed dose to water;Andreo,2000

3. Determination of consensus k Q values for megavoltage photon beams for the update of IAEA TRS-398;Andreo;Phys. Med. Biol.,2020

4. Consistency in reference radiotherapy dosimetry: Resolution of an apparent conundrum when 60Co is the reference quality for charged-particle and photon beams;Andreo;Phys. Med. Biol.,2013

5. Monte Carlo calculation of beam quality correction factors in proton beams using TOPAS/GEANT4;Baumann;Phys. Med. Biol.,2020

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3