Evaluation of Monte Carlo to support commissioning of the treatment planning system of new pencil beam scanning proton therapy facilities

Author:

Botnariuc Daniela,Court Steven,Lourenço Ana,Gosling Andrew,Royle Gary J,Hussein Mohammad,Rompokos Vasilis,Veiga Catarina

Abstract

Abstract Objective: To demonstrate the potential of Monte Carlo (MC) to support the resource-intensive measurements that comprise the commissioning of the treatment planning system (TPS) of new proton therapy facilities.

Approach: Beam models of a pencil beam scanning system (Varian ProBeam) were developed in GATE (v8.2), Eclipse proton convolution superposition algorithm (v16.1, Varian Medical Systems) and RayStation MC (v12.0.100.0, RaySearch Laboratories), using the beam commissioning data. All models were first benchmarked against the same commissioning data and validated on seven spread-out Bragg peak (SOBP) plans. Then, we explored the use of MC to optimise dose calculation parameters, fully explore the performance and limitations of TPS in homogeneous fields and support the development of patient-specific quality assurance (PSQA) processes. We compared the dose calculations of the TPSs against measurements (DDTPSvs.Meas.) or GATE (DDTPSvs.GATE) for an extensive set of plans of varying complexity. This included homogeneous plans with varying field-size, range, width, and range-shifters (RSs) (n=46) and PSQA plans for different anatomical sites (n=11).

Results: The three beam models showed good agreement against the commissioning data, and dose differences of 3.5% and 5% were found for SOBP plans without and with RSs, respectively. DDTPSvs.Meas. and DDTPSvs.GATE were correlated in most scenarios – for example, in homogeneous fields the Pearson’s correlation coefficient was 0.92 and 0.68 for Eclipse and RayStation, respectively. The standard deviation of the differences between GATE and measurements (±0.5% for homogeneous and ±0.8% for PSQA plans) was applied as tolerance when comparing TPSs with GATE. 72% and 60% of the plans were within the GATE predicted dose difference for both TPSs, for homogeneous and PSQA cases, respectively.

Significance: Developing and validating a MC beam model early on into the commissioning of new proton therapy facilities can support the validation of the TPS and facilitate comprehensive investigation of its capabilities and limitations.

Funder

Royal Academy of Engineering

Fundação para a Ciência e a Tecnologia

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3