CDMAM in digital mammography: contrast threshold measurements, achievable values and relation with dose

Author:

Trindade HugoORCID,Barreira Marisa

Abstract

Abstract The main goal of this study is to characterise threshold contrast values and average glandular dose (AGD) in mammography examinations, as well as their relationship with acceptable (minimum quality values) and achievable (desirable quality values) values from the fourth edition of the EUREF protocol. Therefore, data from 84 mammography units from six different manufacturers comprising 15 different models were gathered across Portugal during 2020 and 2021. The contrast measurements were obtained with a CDMAM 4.0 phantom. Within the tested units, six failed to comply with the acceptable EUREF values for threshold contrast and 5% achieved all levels of desirable contrast. Only 3% of the units presented values above an achievable AGD level of 2.4 mGy for an equivalent breast thickness of 60 mm. As these acceptable and achievable values are used as a guideline for the setup of dose and image quality of mammography units, the obtained results demonstrate that clinical images are being produced with an AGD much lower than the achievable dose values. As a result, contrast threshold values are close to the minimum quality values, which could result in images with less diagnostic data, compromising the justification of x-ray exams. Results proved that there is a possibility for an increase in AGD to give achievable image standards. Higher levels of radiation dose correlated with better image quality, and consequently higher diagnostic confidence could be consistent with the optimisation principle for radiation protection.

Publisher

IOP Publishing

Subject

Public Health, Environmental and Occupational Health,Waste Management and Disposal,General Medicine

Reference19 articles.

1. Impact of screening on breast cancer mortality: the UK program 20 years on;Massat;Cancer Epidemiol. Biomarkers Prev.,2016

2. The 2007 recommendations of the International Commission on Radiological Protection. ICRP publication 103;Ann. ICRP,2007

3. Screening mammography: clinical image quality and the risk of interval breast cancer;Taplin;AJR Am. J. Roentgenol.,2002

4. Digital mammography update. European protocol for the quality control of the physical and technical aspects of mammography screening. S1, Part 1: Acceptance and constancy testing;van Engen,2013

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. The human dimension: the Bernard Wheatley Award for 2023;Journal of Radiological Protection;2024-05-16

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3