Abstract
Abstract
The Monte Carlo radiation transport code MCNP6 has been used to model dosimetry for biological pathogen samples placed within a MultiRad 225 irradiation chamber, in order to inform virus deactivation protocols. Full characterisations of the photon spectra generated by the chamber’s x-ray tube were achieved for both 190 and 220 kV potentials, with and without aluminium and copper beam filters of different thicknesses. Dose rate maps to air and water within the chamber were then derived, along with corresponding conversion coefficient data. The maps were determined for samples located both on a shelf and on a dry ice refrigeration chamber, at different distances from the source. The potential depth-dose profiles through samples were also investigated. The optimum choice of filter for use in virus inactivation procedures will rely on a compromise between dose homogeneity and dose rate.
Subject
Public Health, Environmental and Occupational Health,Waste Management and Disposal,General Medicine
Reference9 articles.
1. X-ray inactivation of RNA viruses without loss of biological characteristics;Afrough;Nat. Sci. Rep.,2020
2. Alternatives to caesium irradiators for biological sciences and blood transfusion services;Barnard,2020
3. Inactivation of Lassa, Marburg, and Ebola viruses by gamma irradiation;Elliott;J. Clin. Microbiol.,1982
4. Initial MCNP6 release overview;Goorley;Nucl. Technol.,2012
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献