Impact of shield location on staff and caregiver dose rates for I-131 radiopharmaceutical therapy patients

Author:

Bellamy MORCID,Chu B,Serencsits B,Quinn B,Prasad K,Altamirano J,Williamson M,Miodownik D,Abrahams N,Chen F,Bierman D,Wutkowski M,Carter LORCID,Dauer LORCID

Abstract

Abstract The goal of this study is to investigate the effect of the location and width of a single lead shield on the dose rate of staff and caregivers in a hospital room with an I-131 patient. The best orientation of the patient and caregiver relative to the shield was determined based on minimizing staff and caregiver radiation dose rates. Shielded and unshielded dose rates were simulated using a Monte Carlo computer simulation and validated using real-world ionisation chamber measurements. Based on a radiation transport analysis using an adult voxel phantom published by the International Commission on Radiological Protection, placing the shield near the caregiver yielded the lowest dose rates. However, this strategy reduced the dose rate in only a tiny area of the room. Furthermore, positioning the shield near the patient in the caudal direction provided a modest dose rate reduction while shielding a large room area. Finally, increased shield width was associated with decreasing dose rates, but only a four-fold dose-rate reduction was observed for standard width shields. The recommendations of this case study may be considered as potential candidate room configurations where radiation dose rates are minimized, however these findings must be weighed against additional clinical, safety, and comfort considerations.

Funder

National Institutes of Health/National Cancer Institute

Cancer Center

Publisher

IOP Publishing

Subject

Public Health, Environmental and Occupational Health,Waste Management and Disposal,General Medicine

Reference31 articles.

1. Medical radiation exposure of patients in the United States,2019

2. Radiological protection in medicine. ICRP publication 105;Ann. ICRP,2008

3. Management of radionuclide therapy patients,2006

4. The 2007 recommendations of the International Commission on Radiological Protection. ICRP publication 103;Ann. ICRP,2007

5. Estimated dose rates to members of the public from external exposure to patients with 131I thyroid treatment;Dewji;Med. Phys.,2015

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3