Abstract
Abstract
Dual-energy contrast-enhanced spectral mammography (DE-CESM) is a recently developed advanced technique in digital mammography that uses an iodinated intravenous contrast agent to assess tumor angiogenesis. The aim of this study was to investigate the diagnostic potential of DE-CESM recombined images in terms of radiation dose and image quality. A 50% fibroglandular—50% adipose, custom-made phantom with iodine inserts of 1.0 mgI cm−3, 2.0 mgI cm−3, 4.0 mgI cm−3 was used for the estimation of mean glandular dose (MGD) and the image quality. Low-energy (LE) images were acquired with the W/Rh, W/Rh + 0.01 mm Cu and W/Rh + 0.5 mm Al while high energy images (HE) are acquired with the W/Rh, W/Rh + 0.06 mm Ba, W/Rh + 0.01 mm Cu, and W/Rh + 0.03 mm Ce anode filter combinations. The total MGD was reduced up to a maximum from 1.75 mGy to 1.45 mGy by using Rh + 0.01 mm Cu double-layer filter for both LE and HE imaging of 50 mm, standard 50% fibroglandular phantom compared to Rh single-layer filter with W target. The minimum total MGD reduction (1.69 mGy) was observed when Rh + 0.5 mm Al was used for LE and Rh + 0.06 mm Ba was used for HE exposure. The image quality was comparable with the single-layer filter. The use of W/Rh + 0.01 mm Cu or W/Rh + 0.5 mm Al as target/filter combination for LE exposure and W/Rh + 0.01 mm Cu for HE exposure can reduce the additional radiation dose delivered by DE-CESM without degrading the image quality.