Non-contact quantification of aortic stenosis and mitral regurgitation using carotid waveforms from skin displacements

Author:

Khwaounjoo PrashannaORCID,Dixon Alexander WORCID,HajiRassouliha Amir,Lam Po Tang Emily J,Webster Mark W I,Taberner Andrew JORCID,Nielsen Poul M FORCID,Nash Martyn PORCID,Cakmak Yusuf OORCID

Abstract

Abstract Objective. Early diagnosis of heart problems is essential for improving patient prognosis. Approach. We created a non-contact imaging system that calculates the vessel-induced deformation of the skin to estimate the carotid artery pressure displacement waveforms. We present a clinical study of the system in patients (n = 27) with no underlying condition, aortic stenosis (AS), or mitral regurgitation (MR). Main results. Displacement waveforms were compared to aortic catheter pressures in the same patients. The morphologies of the pressure and displacement waveforms were found to be similar, and pulse wave analysis metrics, such as our modified reflection indices (RI) and waveform duration proportions, showed no significant differences. Compared with the control group, AS patients displayed a greater proportion of time to peak (p = 0.026 and p = 0.047 for catheter and displacement, respectively), whereas augmentation index (AIx) was greater for the displacement waveform only (p = 0.030). The modified RI for MR (p = 0.047 and p = 0.004 for catheter and displacement, respectively) was lower than in the controls. AS and MR were also significantly different for the proportion of time to peak (p = 0.018 for the catheter measurements), RI (p = 0.045 and p = 0.002 for the catheter and displacement, respectively), and AIx (p = 0.005 for the displacement waveform). Significance. These findings demonstrate the ability of our system to provide insights into cardiac conditions and support further development as a diagnostic/telehealth-based screening tool.

Funder

MedTech CoRE

Publisher

IOP Publishing

Subject

Physiology (medical),Biomedical Engineering,Physiology,Biophysics

Reference51 articles.

1. Non-contact hemodynamic imaging reveals the jugular venous pulse waveform;Amelard;Sci. Rep.,2017

2. An accelerometric sensor system with integrated hydrostatic pressure correction to assess carotid arterial stiffness;Arathy;IEEE Sens. J.,2021

3. Arterial stiffness in aortic stenosis: relationship with severity and echocardiographic procedures response;Bruschi;High Blood Press. Cardiovascular Prevention,2017

4. Aortic stenosis: insights on pathogenesis and clinical implications;Carità;J. Geriatr. Cardiol: JGC,2016

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献