Accuracy enhancement in reflective pulse oximetry by considering wavelength-dependent pathlengths

Author:

Badiola IdoiaORCID,Blazek Vladimir,Jagadeesh Kumar V,George Boby,Leonhardt Steffen,Hoog Antink ChristophORCID

Abstract

Abstract Objective. Noninvasive measurement of oxygen saturation (SpO 2) using transmissive photoplethysmography (tPPG) is clinically accepted and widely employed. However, reflective photoplethysmography (rPPG)—currently present in smartwatches—has not become equally accepted, partially because the pathlengths of the red and infrared PPGs are patient-dependent. Thus, even the most popular ‘Ratio of Modulation’ (R) method requires patient-dependent calibration to reduce the errors in the measurement of SpO 2 using rPPGs. Approach. In this paper, a correction factor or ‘pathlength ratio’ β is introduced in an existing calibration-free algorithm that compensates the patient-dependent pathlength variations, and improved accuracy is obtained in the measurement of SpO 2 using rPPGs. The proposed pathlength ratio β is derived through the analytical model of a rPPG signal. Using the new expression and data obtained from a human hypoxia study wherein arterial oxygen saturation values acquired through Blood Gas Analysis were employed as a reference, β is determined. Main results. The results of the analysis show that a specific combination of the β and the measurements on the pulsating part of the natural logarithm of the red and infrared PPG signals yields a reduced root-mean-square error (RMSE). It is shown that the average RMSE in measuring SpO 2 values reduces to 1 %. Significance. The human hypoxia study data used for this work, obtained in a previous study, covers SpO 2 values in the range from 70 % to 100 %, and thus shows that the pathlength ratio β proposed here works well in the range of clinical interest. This work demonstrates that the calibration-free method applicable for transmission type PPGs can be extended to determine SpO 2 using reflective PPGs with the incorporation of the correction factor β. Our algorithm significantly reduces the number of parameters needed for the estimation, while keeping the RMSE below the clinically accepted 2 %.

Funder

Bundesministerium für Wirtschaft und Energie

Publisher

IOP Publishing

Subject

Physiology (medical),Biomedical Engineering,Physiology,Biophysics

Reference27 articles.

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3