Abstract
Abstract
Objective: Accurate and reliable blood pressure (BP) measurement is important for the prevention and treatment of hypertension. The oscillometric-based automatic office blood pressure measurement (AOBPM) is widely used in hospitals and clinics, but measurement errors are common in BP measurements. There is a lack of systematic review of the sources of measurement errors. Approach: A systematic review of all existing research on sources of AOBPM errors. A search strategy was designed in six online databases, and all the literature published before October 2021 was selected. Those studies that used the AOBPM device to measure BP from the upper arm of subjects were included. Main results: A total of 1365 studies were screened, and 224 studies were included in this final review. They investigated 22 common error sources with clinical AOBPM. Regarding the causes of BP errors, this review divided them into the following categories: the activities before measurement, patient’s factors, measurement environment, measurement procedure, and device settings. 13 sources caused increased systolic and diastolic BP (SBP and DBP), 2 sources caused the decrease in SBP and DBP, only 1 source had no significant effect on BPs, and the other errors had a non-uniform effect (either increase or decrease in BPs). The error ranges for SBP and DBP were −14 to 33 mmHg and −6 to 19 mmHg, respectively. Significance: The measurement accuracy of AOBPM is susceptible to the influence of measurement factors. Interpreting BP readings need to be treated with caution in clinical measurements. This review made comprehensive evidence for the need for standardized BP measurements and provided guidance for clinical practitioners when measuring BP with AOBPM devices.
Funder
National Key Research and Development Program of China
National Natural Science Foundation of China
Natural Science Foundation of Jiangsu Province
Subject
Physiology (medical),Biomedical Engineering,Physiology,Biophysics