Author:
Zeng 曾 Mengyan 孟彦,Ma 马 Zixiao 子晓,Hu 胡 Ruming 如明,Zhang 张 Baolin 宝林,Hao 郝 Yanmei 艳梅,Zhang 张 Huaqing 华青,Huang 黄 Yao 垚,Guan 管 Hua 桦,Gao 高 Kelin 克林
Abstract
Future applications of portable 40Ca+ optical clocks require reliable magnetic field stabilization to improve frequency stability, which can be achieved by implementing an active and passive magnetic field noise suppression system. On the one hand, we have optimized the magnetic shielding performance of the portable optical clock by reducing its apertures and optimizing its geometry; on the other hand, we have introduced an active magnetic field noise suppression system to further suppress the magnetic field noise experienced by the ions. These efforts reduced the ambient magnetic field noise by about 10000 times, significantly reduced the linewidth of the clock transition spectrum, improved the stability of the portable 40Ca+ optical clock, and created the conditions for using portable optical clocks in non-laboratory magnetic field environments. This active magnetic field suppression scheme has the advantages of simple installation and wide applicability.