Author:
Li Zhe,Yang Shuang,Zhang Zhirong,Xia Hua,Pang Tao,Wu Bian,Sun Pengshuai,Wang Huadong,Yu Runqing
Abstract
Continuous-wave cavity ring-down spectroscopy (CW-CRDS) is an important technical means to monitor greenhouse gases in atmospheric environment. In this paper, a CW-CRDS system is built to meet the needs of atmospheric methane monitoring. The problem of mode matching is explained from the perspective of transverse mode and longitudinal mode, and the influence of laser injection efficiency on measurement precision is further analyzed. The results of cavity ring-down time measurement show that the measurement precision is higher when the laser is coupled with the fundamental mode. In the experiment, DFB laser is used to calibrate the system with standard methane concentration, and the measurement residual is less than ±4 × 10−4 μs−1. The methane concentration in the air is monitored in real time for two days. The results show the consistency of the concentration changes over the two days, which further demonstrates the reliability of the system for the measurement of trace methane. By analyzing the influence of mode matching, it not only assists the adjustment of the optical path, but also further improves the sensitivity of the system measurement.
Subject
General Physics and Astronomy
Cited by
3 articles.