Author:
Liu Dao-Xin,Cao Jian,Yuan Jin-Bo,Cui Kai-Feng,Yuan Yi,Zhang Ping,Chao Si-Jia,Shu Hua-Lin,Huang Xue-Ren
Abstract
We report an experimental demonstration of geopotential difference measurement using a pair of transportable 40Ca+ optical clocks (TOC-729-1 and TOC-729-3) in the laboratory, each of them has an uncertainty of 1.3 × 10−17 and an instability of
4.8
×
10
−
15
/
τ
. Referenced to a stationary clock of TOC-729-1, the geopotential difference measurements are realized by moving TOC-729-3 to three different locations and the relevant altitude differences are measured with uncertainties at the level of 20 cm. After correcting the systematic shifts (including gravitational red shift), the two-clock frequency difference is measured to be –0.7(2.2) × 10−17, considering both the statistic (1.0 × 10−17) and the systematic (1.9 × 10−17) uncertainties. The frequency difference between these two clocks is within their respective uncertainties, verifying the reliability of transportable 40Ca+ optical clocks at the low level of 10−17.