Abstract
From the perspective of error compensation in the sampling process, a digital calibration algorithm was studied for the processing of spectral data in dual-comb spectroscopy. In this algorithm, dynamic adaptation to phase fluctuations maintained constant measurement results of spectral line positions and intensities. A mode-resolved broadband absorption spectrum was obtained over the full-spectral range of the comb with a Hertz linewidth of radio frequency comb mode. The measured spectrum spanned over 10 THz, which covered the multiplexed absorption regions of mixed gases, such as CO2 and N2O. The calibrated interferograms were also capable of direct coherent averaging in the time domain. The transmittance obtained deviated from the theoretical calculation by no more than 2% in the whole spectral span.