Abstract
Abstract
This paper describes the NRC measurement set-up and the preparatory work for the CCT-K7.2021 key comparison (KC) of triple-point-of-water (TPW) cells. The preparatory work at NRC (charged with piloting the KC) included a comprehensive testing of the measurement set-up, the definition of the new NRC national standard for the TPW temperature, and the evaluation of the uncertainty budgets relevant for this KC. The new NRC TPW national standard is composed of an ensemble of ten fused silica TPW cells, characterized in both their isotopic composition and impurity content. Three uncertainty budgets were evaluated, corresponding to three different cases involved in the measurements at the pilot laboratory: (a) the temperature difference between two TPW cells, which amounted to 26 μK (k = 2), (b) the temperature difference between a TPW cell and NRC national reference, which amounted to 35 μK (k = 2) and (c) the temperature difference between a comparison participant’s transfer cell and the comparison reference, composed of the average of two NRC national reference cells, which amounted to 19 μK (k = 2).
Reference10 articles.
1. The International Temperature Scale of 1990 (ITS-90);Preston-Thomas;Metrologia,1990
2. Mutual recognition of national measurement standards and of calibration and measurement certificates issued by National Metrology Institutes,1999
3. Final report on CCT-K7: key comparison of water triple point cells;Stock;Metrologia,2006
4. A comparison of triple point of water cells;Hill,1997
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献