Abstract
Abstract
We have built an ytterbium optical lattice clock with improvements over our previous version. An in-vacuum blackbody radiation (BBR) shield is employed to provide a well characterized BBR environment. The effective temperature felt by the atoms can be determined at an accuracy level of 13 mK, leading to a total BBR frequency shift uncertainty of 9.5 × 10−19. We have also built an ultra-stable optical cavity system to pre-stabilize the clock laser, achieving a flicker frequency instability of ∼3 × 10−16. Rabi spectroscopy of the lattice-trapped atoms can achieve sub-Hertz linewidth spectra. Two ytterbium clocks have been operated in an antisynchronized configuration, with real-time BBR-Stark-shift corrections applied to both of them. By comparing the two clocks, we demonstrate a single-clock instability of 5.4 × 10−18 in 4500 s. This clock will be applied for frequency comparisons to other optical clocks of different atomic species in the future.
Funder
Strategic Priority Research Program of the Chinese Academy of Sciences
National Natural Science Foundation of China
National Key Research and Development Program of China
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献