Abstract
Abstract
Several correction factors for primary beta dosimetry in radiation protection have been determined by means of measurements and simulations for both the well-known Böhm primary standard extrapolation chamber developed at PTB (which has been commercially available for a long time) and a new primary standard extrapolation chamber also developed at PTB. The correction factors k
el for electrostatic attraction of the entrance window due to the collecting voltage and k
sat for ionisation collection losses due to ionic recombination have been measured. The following correction factors have been determined by simulations for normal (0°) as well as oblique (15°, 30°, 45° and 60°) radiation incidence: k
ba for backscatter from the collecting electrode, k
pe for perturbation by the chamber’s side walls, and k
ra, k
ac and k
di for inhomogeneity inside the collecting volume. The latter three have been combined to k
ih. Six more correction factors adopted from the literature are described to complete the necessary corrections. A comparison with values given in ISO 6980-2:2004 (for 0°), all determined at that time by means of measurements, is shown as well as the application of the new factors to measured data. The newly determined correction factors are ready to be implemented in an updated version of ISO 6980-2.
Reference27 articles.
1. Reference beta-particle radiation—part 1: methods of production,2006
2. Reference beta-particle radiation—part 2: calibration fundamentals related to basic quantities characterizing the radiation field,2004
3. Reference beta-particle radiation—part 3: calibration of area and personal dosemeters and the determination of their response as a function of beta radiation energy and angle of incidence,2006
4. The PTB Beta Secondary Standard BSS 2 for radiation protection JINST;Ambrosi,2007
5. Extensions to the PTB Beta Secondary Standard BSS 2;Behrens,2011