Abstract
Abstract
A well-characterised sample of silicon tunable-barrier electron pump has been operated at a frequency of 2 GHz using a custom drive waveform, generating a pump current of 320 pA. Precision measurements of the current were made as a function of pump control parameters, using a blind protocol, over a 7 week campaign. The combined standard uncertainty for each
∼
10
h measurement was in the range
0.16
−
0.18
parts per million (ppm). The pump current exhibits a plateau along the exit gate voltage flat to approximately 0.15 ppm, but offset from ef by 0.2 ppm. This offset may be a sign of errors in the current traceability chain, indicating a limit to the accuracy of small current scaling using existing methods based on cryogenic current comparators.
Funder
Department for Business, Energy and Industrial Strategy, UK Government