Abstract
Abstract
A type-A uncertainty analysis of the magnitude of the local gravitational field can be used to validate type-B uncertainty estimates of calibrations of triaxial accelerometers by rotation in the gravitational field. We demonstrate the application of this method to validate a type-B uncertainty estimate at the sub 0.1% level of the errors introduced by a high-accuracy, two-axis, accelerometer-characterization system.
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献