Phase transition behavior of pure Fe cells for thermocouple calibration

Author:

Kim Yong-GyooORCID,Kwon Suyong,Kim Sunghun

Abstract

Abstract The fixed points adopted in the international temperature scale of 1990 have generally been used to calibrate contact thermometers, such as thermocouples, to get the highest accuracy. The fixed point for Pd is located at the upper limit range of the Pt/Rh noble metal thermocouple calibration. However, the high price of Pd itself restricts its wide use in calibration laboratories. To overcome this difficulty, this study develops an alternative fixed point having a close transition temperature to the Pd freezing temperature of 1554.8 °C. The alternative is pure Fe with a quoted melting temperature of 1538 °C. To make a thermometric cell, an alumina crucible was used to contain the molten liquid and it was found to be robust enough to last for more than 60 cycles of melting and freezing tests without any chemical and mechanical damage. Two cells were fabricated, and their melting and freezing behaviors were studied using type B thermocouples. By considering the advantage and disadvantages of melting and freezing, the melting temperature was selected as a thermometric transition point for the calibration of thermocouples because of its easy operational processes and acceptable uncertainty level. Using a reference thermocouple that had been calibrated versus a standard radiation thermometer, the average melting temperature of the two Fe cells was 1534.0 °C with an uncertainty of 1.2 °C (k = 2). From the studies, an Fe cell made in an alumina crucible can successfully replace the Pd fixed point.

Publisher

IOP Publishing

Subject

General Engineering

Reference13 articles.

1. The international temperature scale of 1990 (ITS-90);Preston-Thomas;Metrologia,1990

2. Recommended values of temperature on the international temperature scale of 1990 for a selected set of secondary reference points;Bedford;Metrologia,1996

3. Guide on secondary thermometry, specialized fixed points above 0 °C,2018

4. A nickel freezing-point cell for thermocouple calibration;Kim;Metrologia,2001

5. The mini-coil method for calibration of thermocouples at the palladium point;Jahan,2003

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3