Abstract
Abstract
As optical clocks are improved to reach the frequency uncertainty below the 10−17 level, the frequency shift due to the blackbody radiation (BBR) has been one of the major systematic effects hindering further improvement. To evaluate the BBR shift of an Yb optical lattice clock at KRISS, we installed an in-vacuum BBR shield and made radiation thermometry using a black-coated-sphere thermal probe. After we quantitatively measured the conduction loss of the thermal probe and the effects of all the external radiation sources, we determined the temperature at the atom trap site with an uncertainty of 13 mK, which corresponds to an uncertainty of 0.22 mHz in the clock frequency (a fractional frequency of 4.2 × 10−19). The total uncertainty of the BBR shift including the atomic response is 9.5 × 10−19.
Funder
Research on Measurement Standards for Redefinition of SI Units funded by Korea Research Institute of Standards and Science
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献