Senden R J,De Jean P,McAuley K B,Schreiner L J
Abstract
In this work, three new polymer gel dosimeter recipes were investigated that may be more suitable for widespread applications than polyacrylamide gel dosimeters, since the extremely toxic acrylamide has been replaced with the less harmful monomers N-isopropylacrylamide (NIPAM), diacetone acrylamide and N-vinylformamide. The new gel dosimeters studied contained gelatin (5 wt%), monomer (3 wt%), N,N′-methylene-bis-acrylamide crosslinker (3 wt%) and tetrakis (hydroxymethyl) phosphonium chloride antioxidant (10 mM). The NMR response (R
2) of the dosimeters was analysed for conditions of varying dose, dose rate, time post-irradiation, and temperature during irradiation and scanning. It was shown that the dose–response behaviour of the NIPAM/Bis gel dosimeter is comparable to that of normoxic polyacrylamide gel (PAGAT) in terms of high dose-sensitivity and low dependence on dose rate and irradiation temperature, within the ranges considered. The dose–response (R
2) of NIPAM/Bis appears to be linear over a greater dose range than the PAGAT gel dosimeter. The effects of time post-irradiation (temporal instability) and temperature during NMR scanning on the R
2 response were more significant for NIPAM/Bis dosimeters. Diacetone acrylamide and N-vinylformamide gel dosimeters possessed considerably lower dose-sensitivities. The optical dose–response, measured in terms of the attenuation coefficient for each polymer gel dosimeter, showed potential for the use of optical imaging techniques in future studies.
Subject
Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology
Reference48 articles.
1. An NMR relaxometry and gravimetric study of gelatin-free aqueous polyacrylamide dosimeters;Phys. Med. Biol.,2006
2. Experimental procedure for the manufacture and calibration of polyacrylamide gel (PAG) for magnetic resonance imaging (MRI) radiation dosimetry;Phys. Med. Biol.,1998
3. Dose resolution in radiotherapy polymer gel dosimetry: effect of echo spacing in MRI pulse sequence;Phys. Med. Biol.,2001
4. Relaxation effects in nuclear magnetic resonance absorption;Phys. Rev.,1948
5. Experimental study of attenuation properties of normoxic polymer gel dosimeters;Phys. Med. Biol.,2004
Cited by
183 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献