Author:
Lalancette M,Quraan M,Cheyne D
Abstract
Magnetoencephalography (MEG) source analysis has largely relied on spherical conductor models of the head to simplify forward calculations of the brain's magnetic field. Multiple- (or overlapping, local) sphere models, where an optimal sphere is selected for each sensor, are considered an improvement over single-sphere models and are computationally simpler than realistic models. However, there is limited information available regarding the different methods used to generate these models and their relative accuracy. We describe a variety of single- and multiple-sphere fitting approaches, including a novel method that attempts to minimize the field error. An accurate boundary element method simulation was used to evaluate the relative field measurement error (12% on average) and dipole fit localization bias (3.5 mm) of each model over the entire brain. All spherical models can contribute in the order of 1 cm to the localization bias in regions of the head that depart significantly from a sphere (inferior frontal and temporal). These spherical approximation errors can give rise to larger localization differences when all modeling effects are taken into account and with more complex source configurations or other inverse techniques, as shown with a beamformer example. Results differed noticeably depending on the source location, making it difficult to recommend a fitting method that performs best in general. Given these limitations, it may be advisable to expand the use of realistic head models.
Subject
Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology
Cited by
49 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献