Abstract
Abstract
Large deflection sensing is highly crucial for proper positioning and control of continuum robots during robot assisted minimally invasive surgery (MIS). Existing techniques suffer from eletromagnetic noise susceptibility, harmful radiation exposure, limited range, bio-incompatibility and necessity of expensive instruments. In the present study, we propose a multi-walled carbon nanotubes/polyglycerol based low cost, flexible and biocompatible sensor which could allow safer, faster and accurate angular deflection measurement of continuum robots for biomedical applications. Experimental results demonstrate that the sensor is stretchable up to 100%, provides a gauge factor up to 11.65, have response time around 8 ms, durability of −0.14% for cyclic loading and unloading and show very small creep up to
±
0.0008
(
±
2.88
%
)
. Furthermore, the sensor can measure continuum robot deflection up to
±
150
∘
with a sensitivity of 666.67 ohms/degree, with a maximum error of 1.67% and maximum hysteresis of 1.41%. Thus, wide range, low cost, fast response, and biocompatibility justify the potential of the proposed sensor for large deflection sensing of continuum robots during robot assisted MIS.
Subject
Applied Mathematics,Instrumentation,Engineering (miscellaneous)
Reference49 articles.
1. The role of technology in minimally invasive surgery: state of the art, recent developments and future directions;Tonutti;Postgrad. Med. J.,2017
2. Steerable catheters for minimally invasive surgery: a review and future directions;Xiaohua;Comput. Assist. Surg.,2018
3. Continuum robots for medical applications: a survey;Burgner-Kahrs;IEEE Trans. Robot.,2015
4. Constrained motion control of multisegment continuum robots for transurethral bladder resection and surveillance;Bajo,2013
5. Continuous shape estimation of continuum robots using x-ray images;Lobaton,2013
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献