Abstract
Abstract
Calibration traceability can be broken where there are significant, unquantified uncertainties. Thermometry in harsh environments using electrical sensors, such as thermocouples or resistance thermometers, can be unpredictably affected by electro-magnetic sources. For this application, phosphor based temperature sensors with fibre-optic connection have been made following two different approaches—phosphor decay time changes and phosphor emission spectral changes—and tested by bombardment with high energy electrons and by measurements in large magnetic fields. Both sets of tests showed good immunity to exposure, and with magnetic field tests significantly better than a thermocouple. Phosphor thermometry therefore has potential to retain traceability in situations where conventional sensors might fail.
Subject
Applied Mathematics,Instrumentation,Engineering (miscellaneous)
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献