Abstract
Aiming at the problem of inaccurate measurement results caused by factors such as gas leakage during the calibration of the standard meter gas flow standard device of the same type and small diameter, this paper proposes a similarity assessment method of single flow point energy consumption mapping based on Fréchet distance (FD). First, through the statistical analysis of the historical energy consumption data of gas flow standard device, a discrete standard energy consumption mapping is formed. Next, the energy consumption mapping to be tested is standardized and the similarity of the energy consumption mapping is determined by the binary matrix accelerated search of the FD between the energy consumption mapping to be tested and the standard energy consumption mapping. Finally, the similarity threshold is calculated by combining the standard deviation of the FD and the reliability of the gas flowmeter calibration process is evaluated accordingly. The test shows that the method in this paper can effectively determine the reliability of the gas flowmeters calibration results.
Funder
Guangdong Basic and Applied Basic Research Foundation
National Natural Science Foundation of China
Technology Project of Guangdong Administration for Market Regulation
Subject
Applied Mathematics,Instrumentation,Engineering (miscellaneous)