Critical implementation issues of excitation signals for embedded wearable bioimpedance spectroscopy systems with limited resources

Author:

Kallel A YORCID,Bouchaala DORCID,Kanoun OORCID

Abstract

Abstract Wideband excitation signals are essential in bioimpedance spectroscopy for measurements in a time ensuring a quasi-stable measurement condition. In particular, for wearable biomedical systems, due to limited system resources, several aspects regarding measurement time, crest factor, slew rate requirements, frequency distribution, amplitude spectrum, and energy efficiency need to be thoroughly investigated. In this paper, we present an investigation of excitation signals, which includes not only the theoretical aspects but also aspects of real implementation on microcontroller-based systems. At a fixed number of samples and sampling rate, we investigate the implementability of signal frequencies and the resulting spectral efficiency. We focus on sources of signal distortion due to timer and amplitude deviations. The results show that for 4096 samples and a sampling frequency of 1 MHz, wideband signals are 2.76 times faster than a stepped frequency sweep. The multisine signal provides a better energy efficiency and has a lower slew rate requirement on hardware (around 0.3 V µs−1), but has a relatively high crest factor, even after optimization. An exemplary investigation of the distortion of the time/frequency and amplitudes following implementation on a standard industrial advanced RISC machines microcontroller has shown that a sampling rate compensation is required to overcome timer inaccuracies. Furthermore, non-return-to-zero binary signals are more sensitive to distortion due to hardware-related issues and have a lower signal-to-distortion-and-noise (SINAD) ratio than 24 dB, which is lower than the multisine signal, having a SINAD of 31 dB.

Funder

Deutscher Akademischer Austauschdienst

Publisher

IOP Publishing

Subject

Applied Mathematics,Instrumentation,Engineering (miscellaneous)

Reference33 articles.

1. Alpha-dispersion in human tissue;Grimnes;J. Phys.: Conf. Ser.,2010

2. Wearable bioimpedance for continuous and context-aware clinical monitoring;Dutt,2020

3. A wrist watch-type cardiovascular monitoring system using concurrent ECG and APW measurement;Lee;J. Semicond. Technol. Sci.,2016

4. Analytical Model for Blood Glucose Detection Using Electrical Impedance Spectroscopy

5. A non-invasive wearable bioimpedance system to wirelessly monitor bladder filling;Reichmuth,2020

Cited by 26 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Enabling Real-Time Impedance Measurements of Operational Superconducting Circuits of Accelerator Magnets;IEEE Transactions on Applied Superconductivity;2024-08

2. High-performance efficient embedded systems for impedance spectroscopy: Challenges and potentials;Electrochimica Acta;2024-07

3. Parameters Estimation of Cole-Cole Bio-Impedance Model with a Minimum Number of Frequencies;2024 IEEE International Instrumentation and Measurement Technology Conference (I2MTC);2024-05-20

4. i-CardiAx: Wearable IoT-Driven System for Early Sepsis Detection Through Long-Term Vital Sign Monitoring;2024 IEEE/ACM Ninth International Conference on Internet-of-Things Design and Implementation (IoTDI);2024-05-13

5. Phase sensitive detection for embedded sensors;Measurement Science and Technology;2024-02-07