Rostocki Aleksander J,Urbański Michał K,Wiśniewski Roland,Wilczyńska Teresa
Abstract
The paper presents the recent progress made in the accuracy improvement of the manganin pressure sensors at the Warsaw University of Technology (WUT) and at the Institute of Nuclear Energy. The efforts of the authors were concentrated on three problems:
(1) The improvement of the temperature characteristics by the design of a manganin sensor with the temperature compensation achieved by the application of two manganin coils with two different temperature characteristics;
(2) The improvement of the calibration process by
- (a) using a high pressure deadweight piston gauge (with the correction of effective area versus pressure dependence, built at WUT),
- (b) measurements of the manganin resistance as a function of the temperature and of the pressure using a precise multimeter and
- (c) application of statistical methods for long series of measurements.
(3) The modification of the thermal characteristic by implantation of bismuth and krypton ions.
Reference14 articles.
1. Absolute measurements in the pressure range up to 30000 kg/cm2;Phys. Rev.,1940
2. Long time relaxation effect of liquid castor oil under high pressure condition;High Pressure Res.,1989
3. A new design of manganin gauge and its metrological characteristics;Metrologia,1994
4. The hydrostatic pressure effect on electro-resistance properties of soldered connections;Archiwum Elektrotechniki,1975
5. XXXVI MKM, Prace Komisji Metrologii Oddziału PAN w Katowicach,2004
Cited by
14 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献