Abstract
Abstract
This study proposed a C4D sensor structure with a novel approach utilizing the complex impedance analysis method, which includes both resistance and reactance analysis. The complex impedance analysis method provides additional information to characterize the electrical properties of solutions and strange objects moving in the fluidic channel, thus enhancing the sensitivity of conventional C4D structure. In addition, the effects of the fluidic properties, object size, excitation frequency, and sensor geometry on the performance of the proposed C4D sensor were investigated. The proposed sensor system used the finite element method analysis to investigate the performance of the proposed method. The simulation results show that the largest resistance change is approximately 27 kΩ while the largest reactance change is about 10 kΩ according to the passage of a 400 μm radius oil droplet over the sensing area in the NaCl solution channel. Furthermore, this work indirectly verifies the possibility and feasibility of the proposed impedance analysis method by utilizing the conventional C4D structure thanks to its advantages and lays a good foundation for future research work.
Funder
Vietnam Ministry of Science and Technology
Scholarship Programme of Vingroup Innovation Foundation
Subject
Computer Science Applications,Mechanics of Materials,Condensed Matter Physics,General Materials Science,Modeling and Simulation
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献