Preliminary dielectrophoresis study: Manipulation of protein albumin and electrical quantification by using cyclic voltammetry technique
Author:
Abdul Nasir Nur Shahira,Deivasigamani Revathy,Abdul Rahim Muhammad Khairulanwar,Mohd Nashruddin Siti Nur Ashakirin,Hamzah Azrul Azlan,M. Razip Wee M. Farhanulhakim,Buyong Muhamad Ramdzan
Abstract
Purpose
The purpose of this paper is to visualize protein manipulation using dielectrophoresis (DEP) as a substantial perspective on being an effective protein analysis and biosensor method as DEP is able to be used as a means for manipulation, fractionation, pre-concentration and separation. This research aims to quantify DEP using an electrochemical technique known as cyclic voltammetry (CV), as albumin is non-visible without any fluorescent probe or dye.
Design/methodology/approach
The principles of DEP were generated by an electric field on tapered DEP microelectrodes. The principle of CV was analysed using different concentrations of albumin on a screen-printed carbon electrode. Using preliminary data from both DEP and CV methods as a future prospect for the integration of both techniques to do electrical quantification of DEP forces.
Findings
The size of the albumin is known to be 0.027 µm. Engineered polystyrene particle of size 0.05 µm was selected to mimic the DEP actuation of albumin. Positive DEP of the sample engineered polystyrene particle was able to be visualized clearly at 10 MHz supplied with 20 Vpp. However, negative DEP was not able to be visualized because of the limitation of the apparatus. However, albumin was not able to be visualized under the fluorescent microscope because of its translucent properties. Thus, a method of electrical quantification known as the CV technique is used. The detection of bovine serum albumin (BSA) using the CV method is successful. As the concentration of BSA increases, the peak current obtained from the voltammogram decreases. The peak current can be an indicator of DEP response as it correlates to the adsorption of the protein onto the electrodes. The importance of the results from both CV and DEP shows that the integration of both techniques is possible.
Originality/value
The integration of both methods could give rise to a new technique with precision to be implemented into the dialyzers used in renal haemodialysis treatment for manipulation and sensing of protein albumin.
Subject
Electrical and Electronic Engineering,Surfaces, Coatings and Films,Condensed Matter Physics,Atomic and Molecular Physics, and Optics,Electronic, Optical and Magnetic Materials
Reference25 articles.
1. A user’s guide for phase separation assays with purified proteins;Journal of Molecular Biology,2018
2. Glomerular filtration barrier assembly: an insight;Postdoc Journal: a Journal of Postdoctoral Research and Postdoctoral Affairs,2013
3. Dielectrophoretic characterization of array type microelectrodes,2014
4. A tapered aluminium microelectrode array for improvement of dielectrophoresis-based particle manipulation;Sensors,2015
5. Dynamic dielectric properties characterization of tapered dielectrophoresis microelectrodes for selective detection and rapid manipulation of cells,2020
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献