Fabrication of a MEMS accelerometer to detect heart bypass surgery complications
Author:
Lowrie Craig,Desmulliez Marc P.Y.,Hoff Lars,Jakob Elle Ole,Fosse Erik
Abstract
PurposeThe purpose of this paper is to review the design and fabrication of a micro‐accelerometer to be used to measure the heart wall motion of patients who have just undergone coronary artery bypass graft (CABG) surgery. The sensor will provide a means of early warning for the medical staff of associated complications with this surgery occurring.Design/methodology/approachA feasibility paper is carried out with the use of commercially available MEMS three‐axis accelerometers. The sensors are used in animal studies during which the sensor is stitched directly to the surface of a pig's heart. A need for smaller sensors is required and these are designed in‐house and fabricated using a MEMS process. The final dimensions of the sensors are 2.5×3.5×1.4 mm in width, length and height, respectively.FindingsThe results of the feasibility studies demonstrate the viability of this type of sensor for heart wall motion measurement. It is possible to detect abnormalities, which can indicate complications associated with CABG. The sensors presented here are fabricated within the tight overall size specifications deemed necessary for this application.Research limitations/implicationsThis paper demonstrates an application of MEMS for implantable medical sensors.Practical implicationsIn the UK, approximately 300,000 people have a heart attack each year. One of the most common surgeries that is used to treat this is CABG. This sensor is to be used by the medical staff in post‐surgery to provide “real‐time” monitoring of the heart and give early warning of regional cardiac ischemia which can save lives and reduce hospital waiting times and costs.Originality/valueThis paper demonstrates an original way of measuring heart wall motion. Results from the feasibility studies have proven that this can provide an invaluable way of providing early warning of complications after heart surgery.
Subject
Electrical and Electronic Engineering,Industrial and Manufacturing Engineering
Reference7 articles.
1. Elle, O.J., Halvorsen, S., Gulbrandsen, M.G. and Fosse, E. (2005), “Early recognition of regional cardiac ischemia using a 3‐axis accelerometer sensor”, Physiological Measurement, Vol. 26 No. 4, pp. 429‐40. 2. Imenes, K., Aasmundtveit, K. and Husa, E.M. (2007), “Assembly and packaging of a three‐axis micro accelerometer used for detection of heart infarction”, Springer Biomedical Microdevices, Vol. 9 No. 6, pp. 951‐7. 3. Imenes, K., Aasmundtveit, K. and Moreno, P. (2008), “Micro ribbon cable bonding for an implantable device”, Proceedings of the 2nd Electronics Systemintegration Technology Conference, 1‐4 September, University of Greenwich, London, pp. 265‐9. 4. NHS (2007), “Coronary heart disease”, available at: www.nhs.uk/conditions/Coronary‐heart‐disease/(accessed 15 January 2009).. 5. Velten, T., He, C.Y. and Obermeie, E. (1997), “Dynamic behavior of a new two‐axis accelerometer”, International Conference on Solid State Sensors and Actuators, TRANSDUCERS '97, Chicago, IL, pp. 1217‐20.
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
|
|