Han Changjun,Yan Chunze,Wen Shifeng,Xu Tian,Li Shuai,Liu Jie,Wei Qingsong,Shi Yusheng
Abstract
Purpose
Selective laser melting (SLM) is an additive manufacturing process suitable for fabricating metal porous scaffolds. The unit cell topology is a significant factor that determines the mechanical property of porous scaffolds. Therefore, the purpose of this paper is to evaluate the effects of unit cell topology on the compression properties of porous Cobalt–chromium (Co-Cr) scaffolds fabricated by SLM using finite element (FE) and experimental measurement methods.
Design/methodology/approach
The Co-Cr alloy porous scaffolds constructed in four different topologies, i.e. cubic close packed (CCP), face-centered cubic (FCC), body-centered cubic (BCC) and spherical hollow cubic (SHC), were designed and fabricated via SLM process. FE simulations and compression tests were performed to evaluate the effects of unit cell topology on the compression properties of SLM-processed porous scaffolds.
Findings
The Mises stress predicted by FE simulations showed that different unit cell topologies resulted in distinct stress distributions on the bearing struts of scaffolds, whereas the unit cell size directly determined the stress value. Comparisons on the stress results for four topologies showed that the FCC unit cell has the minimum stress concentration due to its inclined bearing struts and horizontal arms. Simulations and experiments both indicated that the compression modulus and strengths of FCC, BCC, SHC, CCP scaffolds with the same cell size presented in a descending order. These distinct compression behaviors were correlated with the corresponding mechanics response on bearing struts. Two failure mechanisms, cracking and collapse, were found through the results of compression tests, and the influence of topological designs on the failure was analyzed and discussed. Finally, the cell initial response of the SLM-processed Co-Cr scaffold was tested through the in vitro cell culture experiment.
Originality/value
A focus and concern on the compression properties of SLM-processed porous scaffolds was presented from a new perspective of unit cell topology. It provides some new knowledge to the structure optimization of porous scaffolds for load-bearing bone implants.
Subject
Industrial and Manufacturing Engineering,Mechanical Engineering
Reference38 articles.
1. Reconstructing the human body using biomaterials;JOM,1998
2. Mechanical behavior of regular open-cell porous biomaterials made of diamond lattice unit cells;Journal of the Mechanical Behavior of Biomedical Materials,2014
3. Porosity and the fatigue behavior of hypoeutectic and hypereutectic aluminum-silicon casting alloys;International Journal of Fatigue,2008
4. Corrosion aspects of metallic implants – An overview;Materials and Corrosion,2008
5. Application of laser engineered net shaping (LENS) to manufacture porous and functionally graded structures for load bearing implants;Journal of Materials Science: Materials in Medicine,2009
Cited by
63 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献