Purpose
The purpose of this study is to deal largely with the influence of temperature variation on the measurement accuracy of transit-time ultrasonic flowmeter.
Design/methodology/approach
The causes of measurement error due to temperature are qualitatively and quantitatively analyzed, and a mathematical model is established. The experimental data are processed and analyzed, and the temperature compensation coefficient of flow measurement is obtained.
Findings
The experimental results show that the flow measurement results by temperature compensation are helpful in improving the measurement accuracy of the ultrasonic flowmeter.
Practical implications
This study has certain application value, which can provide theoretical support for the design of high-precision ultrasonic flowmeters and design guidance.
Originality/value
It is worth emphasizing that there are few research studies on the influence factors of temperature. This paper focuses on the influence of the temperature change on the flowmeter that is modeled, and the high precision flow parameter test system is designed based on the established model.
Subject
Electrical and Electronic Engineering,Industrial and Manufacturing Engineering
Reference26 articles.
1. High-frequency transducers and correlation method to enhance ultrasonic gas flow metering;Flow Measurement & Instrumentation,2001
2. Method for improving accuracy of ultrasonic flowmeter by using time difference method;Microelectronics & Computer,2009
3. Water density determination in high-accuracy flowmeter calibration – measurement uncertainties and practical aspects;Flow Measurement & Instrumentation,2012
4. Measurement of velocity in rotational flows using ultrasonic anemometry: the flowmeter;Experiments in Fluids,2007
5. A new calibration facility for water flowrate at high Reynolds number;Flow Measurement & Instrumentation,2009
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献