An evaluation of the characteristics of membrane materials suitable for the batch fabrication of dissolved oxygen sensors
Author:
Glasspool W.V.,Atkinson J.K.
Abstract
The bulk batch fabrication process of thick film technology has been utilised in the design and production of miniature amperometric dissolved oxygen sensors based on potentiostatic and voltammetric operation. Three different polymers have been investigated as membrane materials – cellulose acetate, PTFE and PVC. PTFE has been deposited on the devices by aerosol spray and PVC and cellulose acetate by screen‐printing. These methods have been shown to be effective membrane fabrication techniques, and have significant implications in the field of chemical sensors as a whole. All the membrane covered devices investigated were found to exhibit sensitive and linear responses to dissolved oxygen. The effects of temperature and flow rate on sensor response have been investigated and the use of PVC and PTFE in place of cellulose acetate have been shown to reduce both effects. These membranes have also been shown to reduce the detrimental effects of fouling observed on the surfaces of cellulose acetate covered devices as they are powered in tap water.
Subject
Electrical and Electronic Engineering,Surfaces, Coatings and Films,Condensed Matter Physics,Atomic and Molecular Physics, and Optics,Electronic, Optical and Magnetic Materials
Reference24 articles.
1. Adzic, R.R., Tripkovic, A.V. and Markovic, N.M. (1980), “Oxygen reduction on electrode surfaces modified by foreign metal ad‐atoms: lead ad‐atoms on gold”, J. Electroanal. Chem., Vol. 114, pp. 37–51. 2. Atkinson, J.K., Cranny, A.W.J., Glasspool, W.V. and Mihell, J.A. (1999), “An investigation of the performance characteristics and operational lifetimes of multi‐element thick film sensor arrays used in the determination of water quality parameters”, Sensors and Actuators, B: Chemical, Vol. 54 No. 3, pp. 215–31. 3. Atkinson, J.K., Shahi, S., Varney, M. and Hill, M. (1991), “A thick‐film electrochemical instrument”, Sensors and Actuators, B, Vol. 4, pp. 175–81. 4. Cha, G.S., Liu, D., Meyerhoff, M.E., Cantor, C.C., Midgley, A.R., Goldberg, H.D. and Brown, R.B. (1991), “Electrochemical performance, biocompatibility, and adhesion of new polymer matrices for solid state ion sensors”, Anal. Chem., Vol. 63, pp. 239–46. 5. Chen, Y‐O. and Li, G. (1993), “An autocalibrated miniature microhole cathode array for measuring dissolved oxygen”, Sensors and Actuators, B, Vol. 10, pp. 219–22.
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
|
|