Shaul C,Dragovic A S,Stringer A K,O'Leary S J,Briggs R J
Abstract
AbstractObjectiveTo identify the intracochlear electrode position in cochlear implant recipients and determine the correlation to speech perception for two peri-modiolar electrode arrays.MethodsPost-operative cone-beam computed tomography images of 92 adult recipients of the ‘CI512’ electrode and 18 adult recipients of the ‘CI532’ electrode were analysed. Phonemes scores were recorded pre-implantation, and at 3 and 12 months post-implantation.ResultsAll CI532 electrodes were wholly within scala tympani. Of the 79 CI512 electrodes intended to be in scala tympani, 58 (73 per cent) were in scala tympani, 14 (17 per cent) were translocated and 7 (9 per cent) were wholly in scala vestibuli. Thirteen CI512 electrodes were deliberately inserted into scala vestibuli. Speech perception scores for post-lingual recipients were higher in the scala tympani group (69.1 per cent) compared with the scala vestibuli (54.2 per cent) and translocation (50 per cent) groups (p < 0.05). Electrode location outside of scala tympani independently resulted in a 10.5 per cent decrease in phoneme scores.ConclusionCone-beam computed tomography was valuable for demonstrating electrode position. The rate of scala tympani insertion was higher in CI532 than in CI512 electrodes. Scala vestibuli insertion and translocation were associated with poorer speech perception outcomes.