Author:
Shone G. R.,Griffith I. P.
Abstract
AbstractTitanium grommets have been promoted as having the advantage of a slower extrusion rate than other types of ventilation tube. A prospective trial was therefore designed to compare the function and extrusion rates of these grommets with those of the widely used Shepard design of Teflon grommet in a single group of patients. Thirty-one children had one type of grommet inserted in one ear and the other type in the opposite ear. After eight months there were significantly more Titanium grommets still functioning (p < 0.05) but after 12 and after 16 months there was no significant difference in the extrusion rates of the two types of grommet. There was a higher incidence of infection with granulation tissue formation around the Titanium grommet. Accordingly it is concluded that the extra expense of the Titanium grommet is not justified, particularly as the long-term effects of these grommets on the tympanic membrane are not known.
Publisher
Cambridge University Press (CUP)
Subject
Otorhinolaryngology,General Medicine
Cited by
21 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献