Author:
Tjellström A,Granström G,Odersjö M
Abstract
Objective: The purpose of this study was to investigate whether a new self-tapping implant for a bone-anchored hearing aid (BAHA) had the same high frequency of osseointegration as previous implants requiring pre-tapping.Method: Over a three-year period, 144 consecutive implants were placed in the mastoid for BAHA and evaluated.Results: Two implants were lost; both were of the self-tapping type. One was in an 11-year-old boy, who lost his implant six weeks after surgery when the BAHA was fitted. The other was in an elderly man, a heavy smoker with diabetes. Using Fisher's exact test, there was no significant difference between the two groups (p>0.30).Conclusion: Self-tapping implants facilitate surgery and shorten operating time. Over a short follow up, we did not find any significant difference; however, it is important to follow these implants over a longer time period.
Publisher
Cambridge University Press (CUP)
Subject
Otorhinolaryngology,General Medicine
Cited by
18 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献