Author:
Masciocchi Norberto,Artioli Gilberto
Abstract
A well-crystallized sample of powdered KCl has been distributed among several laboratories in order to test the reproducibility of the lattice parameter measurement on different X-ray powder diffraction instruments. The precision of the determined unit-cell dimension is in the 10−5 Å range, while the discrepancies among the results from different laboratories using the same numerical analysis are at least one order of magnitude higher. It is shown that if different numerical analyses, including full pattern refinement, are used, values differing in the third decimal digit are obtained for the same data set.
Publisher
Cambridge University Press (CUP)
Subject
Condensed Matter Physics,Instrumentation,General Materials Science,Radiation
Cited by
9 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献