Author:
Altaf Saima,Iqbal Khalid,Akram Muhammad,Buzdar Saeed A.
Abstract
AbstractAimThe purpose of this study was to analyse the comparison of intensity-modulated radiation therapy quality assurance (IMRT QA) using Gafchromic® EBT3 film, Electronic portal imaging device (EPID) and MapCHECK®2.BackgroundPretreatment authentication is the main apprehension in advanced radiation therapy treatment plans such as IMRT.Materials and methodsA total of 20 patients were planned on Eclipse treatment planning system using 6 and 15 MV separately.ResultsGamma index of EBT3 film results shows the following average passing rates: 97% for 6 MV and 96·6% for 15 MV using criteria of ±5% of 3 mm, ±3% of 3 mm and ±3% of 2 mm for brain. However, by using ±5% of 3 mm and ±3% of 3 mm criteria, the average passing rates were 95·4% on 6 MV and 95·2% on 15 MV for prostate. For EPID, the results show the average passing rates as 97·8% for 6 MV and 97·2% for 15 MV in for brain. In cases in which ±5% of 3 mm and ±3% of 3 mm were used, the average passing rates were 96·6% for 6 MVand 96·1% for 15 MV for prostate. MapCHECK®2 results show average passing rates of 96·4% for 6 and 96·2% for 15 MV, respectively, for brain using criteria of ±5% of 3 mm, ±3% of 3 mm and ±3% of 2 mm, whereas for ±5% of 3 mm and ±3% of 3 mm the average rates are 95·2% for 6 and 94·7% for 15 MV in prostate.ConclusionsThe EPID results are better than the other methods, and hence EPID can be used effectively for IMRT pretreatment verifications.
Publisher
Cambridge University Press (CUP)
Subject
Oncology,Radiology Nuclear Medicine and imaging
Cited by
1 articles.