Abstract
AbstractAim:In this study, the radiation contamination dose (RCD) for different combinations of electron energy/distance, applicator and radius around the light intraoperative accelerator (LIAC), a high dose per pulse dedicated intraoperative electron radiotherapy machine, has been estimated. Being aware about the amount of RCDs is highly recommended for linear medical electron accelerators.Methods and methods:Monte Carlo Nuclear Particles (MCNP) code was used to simulate the LIAC® head and calculate RCDs. Experimental RCDs measurements were also done by Advanced Markus chamber inside a MP3-XS water phantom. Relative differences of simulations and measurements were calculated.Result:RCD reduction by distance from the machine follows the inverse-square law, as expected. The RCD was decreased by increasing angle from applicator walls opposed to the electron beam direction. The maximum differences between the simulation and measurement results were lower than 3%.Conclusions:The RCD is strongly dependent on electron beam energy, applicator size and distance from the accelerator head. Agreement between the MCNP results and ionometric dosimetry confirms the applicability of this simulation code in modelling the intraoperative electron beam and obtaining the dosimetric parameters. The RCD is a parameter that would restrict working with LIAC in an unshielded operative room.
Publisher
Cambridge University Press (CUP)
Subject
Oncology,Radiology, Nuclear Medicine and imaging
Reference18 articles.
1. 18. ICRU. Radiation dosimetry: electrons with initial energies between 1 and 50 MeV. Report No. 35. Washington, DC: International Commission on Radiation Units and Measurement, 1984.
2. Dosimetric characteristics of electron beams produced by two mobile accelerators, Novac7 and Liac, for intraoperative radiation therapy through Monte Carlo simulation
3. Radiation exposure of personnel during Intraoperative Radiotherapy (IORT): radiation protection aspects;Strigari;J Exp Clin Cancer Res,2004
4. Radiation protection measurements around a 12 MeV mobile dedicated IORT accelerator
5. Intraoperative radiotherapy using a mobile electron LINAC: a retroperitoneal sarcoma case;Sam Beddar;J Appl Clin Med Phys,2005