Author:
Batson P. E.,Chen C. H.,Silcox J.
Abstract
Electron energy loss experiments combined with microscopy have proven to be a valuable tool for the exploration of the structure of electronic excitations in materials. These types of excitations, however, are difficult to measure because of their small intensity. In a usual situation, the filament of the microscope is run at a very high temperature in order to present as much intensity as possible at the specimen. This results in a degradation of the ultimate energy resolution of the instrument due to thermal broadening of the electron beam.We report here observations and measurements on a new LaB filament in a microscope-velocity spectrometer system. We have found that, in general, we may retain a good energy resolution with intensities comparable to or greater than those available with the very high temperature tungsten filament. We have also explored the energy distribution of this filament.
Publisher
Cambridge University Press (CUP)
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Notes and References;Principles of Electron Optics;2018