Author:
Rutala William A.,Gergen Maria F.,Weber David J.
Abstract
Abstract
We evaluated the robustness of sterilization technologies when spores and bacteria were placed on “dirty” instruments and overlaid with blood. The results illustrate that steam sterilization is the most effective sterilization technology with the largest margin of safety, followed by ethylene oxide and hydrogen peroxide gas plasma.
Publisher
Cambridge University Press (CUP)
Subject
Infectious Diseases,Microbiology (medical),Epidemiology
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献